Electronic Equipment > Patient Monitors > BM3 Patient Monitor
BM3 Patient Monitor
Ref: 33718
QTY: 
Add Enquiry
Description
OK Medical Dots

 Heart/pulse rate
 ST analysis, PVC count display
 3 lead ECG (5 lead optional)
 Non-invasive blood pressure (NIBP)
 Nellcor Pulse oximetry (SpO2) - pitch tone
 Impedance respiration
 Temperature (sensor is optional)
 Pacemaker detection indicated on ECG dispay
 Battery and direct DC (18 VDC) input (not for recharge)
 Built-in printer (58 mm)
 120 hours memory trend capacity
 Remote alarm function
 20 alarm memory
 wired or wireless LAN and HL7
 Internal software selectable in GB, DE, FR, IT, PT, PL, RO, HU, ES, CZ, BG, RU, TR, CN.
 User manual: English (on request: Italian, Spanish, French, German)

Technical Data
OK Medical Dots

GENERAL
• Display: 7" colour TFT (800x480)
• Dimensions/Weight: 238x250xh 163 mm/3.1 kg
• Indicators: U p to 4 waves (2xECG, SpO2, Respiration),
• Categorized alarms (3 priority levels), external
• Power LED, heart rate tone, battery status
• Thermal printer: 25, 50 mm/sec (paper width 58 mm)
• Battery: internal battery: Li-ion
• Operating time: 2 hours (fully charged)
• Interfaces: lan digital output for transferring data
• DC input connector: 18 VDC, 2.5 A max
• Operating voltage: 100-240 V, 50-60 Hz

ECG
• Leads: 3 leads, 5 leads(OPTlON)
• Heart Rate Range: 30 to 300 bpm
• Heart Rate Accuracy: ± 3 bpm
• ST level: yes
• PVC Count display: yes
• Bandwidth: 0.5 Hz to 40Hz
• Display Sweep Speeds: 25mm/sec
• ECG Size (Sensitivity): 0.5, 1, 2, 4 mV/cm
• Pacemaker Detection Mode: Indicator on waveform display
• CMRR: >90 dB at 50 or 60Hz
• Input Dynamic Range: ± 5 mVAC, ± 300 mVDC
• Defibrillator Discharge: <4 s
• Arrhythmia analysis: VTAC/VFIB/ ASYSTOLIC
• Lead-Off Detection with display indicator

SpO2
• % Saturation Range: 0% to 100%
• Pulse Rate Range: 0 to 254 bpm
• SpO2 Accuracy: 70% to 100% ±2 digits
• Pulse Rate Accuracy: ± 2 bpm

NIBP
• Technique: O scillometric
• Measurement Modes: Manual: Single measurement
• Auto: Automatic intervals of 1, 2, 3, 4, 5, 10, 15, 20, 30 min. and 1, 2, 4, 8 hours
• Stat: Continuous measurement for 5 minutes
• Cuff pressure display: 50 to 300 mmHg
• B.P. measurement range: Systolic: 0 to 260 mmHg
• Mean Arterial Pressure: 30 to 220 mmHg
• Diastolic: 20 to 200 mmHg
• Adjustable Cuff Inflation Pressure

Respirations
• Range: 5 to 120 breaths/min
• Accuracy: ± 3 breaths/min

Temperature Unit
• Range: 15° to 45°C (59°F to 113°F)
• Accuracy: ± 0.1°C
• Compatible with: Y SI Series 400 Temperature Probes


Related Products


Vital-s Patient Monitor
Temperature Sensor - Skin Type
Vital-Snet Patient Monitor